Click "QT" tab in "Analyze" dialog.
This dialog defines the parameters for the QT-related Analysis. QT-related Analysis is possible only
if the heartbeat average is enabled. QT-related Analysis will always be performed on the interval
from the heartbeat time stamp to the last data point of the average heartbeat. The parameter
"Detection Deadtime" defines a time window after the heartbeat time stamp within which no
analysis will be performed, to avoid detections on artefacts originating from the sodium peak.